DATASCOPE CORP. - FAIRFIELD MEGA 7.5FR. 40CC IAB; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Catalog Number 0684-00-0293
Medical Device Problem Code Inflation Problem (1310)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/16/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Event site postal code: (b)(6).The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id # (b)(4).
 
Event or Problem Description
It was reported that after the insertion of the intra-aortic balloon (iab), the iab was not augmenting and could not inflate.The console generated check iab catheter alarm.There was no patient harm or adverse event reported.
 
Additional Manufacturer Narrative
The device has not been returned to the manufacturer so we are unable to complete an evaluation.Analysis of production (3331/213) - the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported lot/serial number.Based on the review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis (4109/213) - the review of the historical data was performed.Trend analysis (4110/213) - the overall complaint trend data for the period jun-19 through may-21 was reviewed.Communication/interviews: (4111/213) communication/interviews were performed to obtain all possible information.Reference complaint #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEGA 7.5FR. 40CC IAB
Common Device Name
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
MDR Report Key11949442
Report Number2248146-2021-00367
Device Sequence Number11807011
Product Code DSP
UDI-Device Identifier10607567111339
UDI-Public10607567111339
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K120868
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/09/2022
Device Catalogue Number0684-00-0293
Device Lot Number3000110169
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 05/18/2021
Supplement Date Received by Manufacturer08/11/2021
Initial Report FDA Received Date06/07/2021
Supplement Report FDA Received Date08/11/2021
Patient Sequence Number1
Patient Age73 YR
Patient Weight67
-
-