| Model Number |
3L80-22 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/13/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A review of tickets determined that there is normal complaint activity for lot 58505uq11.Trending review determined no trend identified for falsely depressed results for the product.Return testing was not completed as returns were not available.Maintenance of the non-abbott water quality system and re-establishing the required product water quality to the instrument corrected the results issue.Results were returned to expected levels.A review of the manufacturing documentation did not identify any issues associated with the customer's observation.The device history record review identified no non-conformances or deviations related to the complaint issue.A review of the labeling concluded that the issue is sufficiently addressed.Based on the investigation, no systemic issue or deficiency of the architect carbon dioxide reagent, lot number 58505uq11 was identified.
|
| |
|
Event or Problem Description
|
|
The customer reported a high anion gap for a patient while using the architect c4000 analyzer.The following data was provided (normal range anion gap: 11-16): sodium = current = 133 mmol/l / previous = 143 mmol/l.Potassium = current = 4.6 mmol/l / previous = 5.1 mmol/l.Chloride = current = 107 mmol/l / previous = 113 mmol/l.Bicarbonate (co2) = current = 17 mmol/l / previous = 23 mmol/l.Anion gap = current = 18 / previous = 12.The anion gap is calculated measurement from sodium, chloride and bicarbonate (co2).There was no impact to patient management reported.
|
| |
|
Search Alerts/Recalls
|