• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MENTOR TEXAS ARTOURA BREAST TISSUE EXPANDER; EXPANDER, SKIN, INFLATABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS ARTOURA BREAST TISSUE EXPANDER; EXPANDER, SKIN, INFLATABLE Back to Search Results
Model Number TEXP120RUH
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Deformity/ Disfigurement (2360)
Date of Event 05/11/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: right deflation.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a (b)(6) year-old female patient underwent primary breast augmentation with a 535cc artoura breast tissue expander and experienced right side expander deflation postoperatively.As a result, the expander was explanted and replaced with catalog number texp120ruh; serial number (b)(4) on (b)(6) 2021.The impacted serial number is either (b)(4).Supplemental report will be submitted when information is received.
 
Additional Manufacturer Narrative
On june 14, 2021, mentor became aware that the impacted serial number was (b)(6).Field d4 has been updated on this form.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On june 25, 2021, the mentor failure analysis lab received the device for evaluation.The analysis has begun, but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On (b)(6) 2021, device evaluation was completed.Device evaluation summary: investigation was performed to see if root cause of deflation could be determined.The investigation included a review of the manufacturing records for lot-7632601, and a visual evaluation of the device.No leak test was necessary to determine the location of the tear as we observed a tear at the union between the shell and the suture tab in the 6 o¿clock position.Additionally, the shell of the tissue expander was observed blue.When evaluated under a microscope the tear pattern did not reveal parallel striations that are consistent with markings made by a sharp instrument.The results also confirmed that there were no abnormalities with the manufacturing of the tissue expander that would have impacted its performance.Therefore, we were unable to identify the root cause of this tear.In addition, as stated in the product insert data sheet, is contraindicated to place drugs or substances inside the tissue expanders other than sterile saline for injection since this could compromise the product integrity.Potential cause: while deflation is historically the most common complication related to tissue expanders, the rate of deflation with mentor® tissue expanders remains very low.An incident of this nature may be the result of one or more of the following factors: over inflation of the tissue expander, use of the tissue expander in emerging surgical techniques not identified in the instructions for use (ifu), continuous and sustained stresses to the tissue expander, vigorous body movement (e.G.Physical exercise) or excessive manipulation/ trauma in the region of the expander.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTOURA BREAST TISSUE EXPANDER
Common Device Name
EXPANDER, SKIN, INFLATABLE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
MDR Report Key11952480
Report Number1645337-2021-06376
Device Sequence Number1059170
Product Code LCJ
UDI-Device Identifier00081317009542
UDI-Public00081317009542
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
UNK
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 05/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/21/2022
Device Model NumberTEXP120RUH
Device Catalogue NumberTEXP120RUH
Device Lot Number7632601
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/25/2021
Initial Date Received by Manufacturer 05/11/2021
Supplement Date Received by Manufacturer06/14/2021
06/25/2021
07/07/2021
Initial Report FDA Received Date06/07/2021
Supplement Report FDA Received Date06/15/2021
07/06/2021
07/07/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age43 YR
-
-