• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TORNIER INC PERFORM HUMERAL SYSTEM; SHOULDER JOINT METAL/POLYMER SEMI-CONSTRAINED CEMENTED PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TORNIER INC PERFORM HUMERAL SYSTEM; SHOULDER JOINT METAL/POLYMER SEMI-CONSTRAINED CEMENTED PROSTHESIS Back to Search Results
Catalog Number MWM005
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 05/12/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned at the time of this report.If additional information becomes available, it will be provided on a supplemental report.
 
Event or Problem Description
It was reported that one of the spikes from the size 3 cut protector broke off during use in shoulder surgery.The surgeon verified that the broken portion was in the patient and could not remove the portion from the humeral canal.No additional complications were reported.
 
Event or Problem Description
It was reported that one of the spikes from the size 3 cut protector broke off during use in shoulder surgery.The surgeon verified that the broken portion was in the patient and could not remove the portion from the humeral canal.No additional complications were reported.
 
Additional Manufacturer Narrative
The reported event could be confirmed.The device was returned for evaluation and found to be in the condition stated in the event description.The device inspection revealed the following: visual inspection: the received device shows signs of usage evidenced by the scratches and scuffs on the underside of the device.The inspection also confirms separation of one of the spikes that hold the cut protector in place on the resected humeral bone.Based on investigation, the root cause is attributed to wear due to the intended use of the device.The spikes of the device are laser welded on the flat plate.This device is subject to numerous impaction; and pulling forces during it¿s use.If the user used a "joysticking" motion when removing the device from the resected humerus it could likely lead to the breakage noted in this complaint.A review of the device history record for the reported lot did not indicate any abnormalities.A review of the labeling did not indicate any abnormalities.If any further information is provided, the complaint report will be updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERFORM HUMERAL SYSTEM
Common Device Name
SHOULDER JOINT METAL/POLYMER SEMI-CONSTRAINED CEMENTED PROSTHESIS
Manufacturer (Section D)
TORNIER INC
10801 nesbitt avenue s
bloomington MN 55437
Manufacturer (Section G)
TORNIER INC
10801 nesbitt avenue s
bloomington MN 55437
Manufacturer Contact
nathe hendricks
1023 cherry rd
memphis, TN 38117
9014516318
MDR Report Key11952834
Report Number3004983210-2021-00051
Device Sequence Number4491788
Product Code PAO
UDI-Device Identifier00846832082810
UDI-Public00846832082810
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K201315
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberMWM005
Device Lot NumberFF0220287
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/09/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/12/2021
Supplement Date Received by Manufacturer01/05/2022
Initial Report FDA Received Date06/07/2021
Supplement Report FDA Received Date02/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/14/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1