Model Number P154
Medical Device Problem Codes Therapy Delivered to Incorrect Body Area (1508); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/12/1999
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This complaint was found during a recent clinical evaluation review/literature search of this device.The citation is as follows: slonim, s.M., dake, m.D., razavi, m.K., kee, s.T., samuels, s.L., rhee, j.S., & semba, c.P.(1999).Management of misplaced or migrated endovascular stents.Journal of vascular and interventional radiology : jvir, 10(7), 851¿859.Https://doi.Org/10.1016/s1051-0443(99)70127-2.Specific product details are not available.The exact event date is unknown.Complaint conclusion: according to the literature article by slonim, s.M., dake, m.D., razavi, m.K., kee, s.T., samuels, s.L., rhee, j.S., & semba, c.P.(1999).Management of misplaced or migrated endovascular stents.Journal of vascular and interventional radiology : jvir, 10(7), 851¿859.Https://doi.Org/10.1016/s1051-0443(99)70127-2, a palmaz 15mm p154 stent was intended to be placed in the right renal artery but migrated to the aorta due to it dislodging from an unknown balloon.The stent remained on the unknown 0.035¿ guidewire.The stent was repositioned in the left external iliac artery (l eia) percutaneously with another unknown 3-5mm balloon catheter.The product was not returned for analysis.No lot number was provided therefore a product history record (phr) review could not be generated.The reported ¿stent dislodged - in patient¿ and ¿stent inaccurate placement¿ could not be confirmed as the device was not returned for analysis.The exact cause could not be determined.Vessel characteristics are unknown.As no lot number was supplied a phr could not be completed.According to the safety information in the instructions for use ¿the palmaz balloon-expandable peripheral stent is a 316l stainless steel, slotted tube.The stent should be crimped over the recommended balloon delivery catheter.See materials required table for the recommended cordis maxi¿ ds, and powerflex® plus individual product descriptions, and recommended accessory devices.¿ the recommended balloon was the powerflex plus.It is not known if the recommended balloon was used in this case.The limited information available does not suggest a design or manufacturing related cause for the reported event; therefore, no corrective/preventive action will be taken at this time.
 
Event or Problem Description
According to the literature article by slonim, s.M., dake, m.D., razavi, m.K., kee, s.T., samuels, s.L., rhee, j.S., & semba, c.P.(1999).Management of misplaced or migrated endovascular stents.Journal of vascular and interventional radiology : jvir, 10(7), 851¿859.Https://doi.Org/10.1016/s1051-0443(99)70127-2, a palmaz 15mm p154 stent was intended to be placed in the right renal artery but migrated to the aorta due to it dislodging from an unknown balloon.The stent remained on the unknown 0.035¿ guidewire.The stent was repositioned in the left external iliac artery (l eia) percutaneously with another unknown 3-5mm balloon catheter.The device will not be returned for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STENT PER 15MM MEDIUM
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
CORDIS CORPORATION
14201 nw 60 avenue
miami lakes FL 33014
MDR Report Key11954477
Report Number9616099-2021-04612
Device Sequence Number16474198
Product Code FGE
UDI-Device Identifier20705032057247
UDI-Public20705032057247
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source literature
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberP154
Device Catalogue NumberP154
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/25/2021
Initial Report FDA Received Date06/08/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN 3-5MM BALLOON CATHETER; UNKNOWN BALLOON CATHETER
Outcome Attributed to Adverse Event Life Threatening; Required Intervention;
-
-