• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC SOFAMOR DANEK USA, INC TRIMLINE; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC SOFAMOR DANEK USA, INC TRIMLINE; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 8756481
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/17/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Evaluation of the returned device was not completed at the time of this report.A follow up report will be sent when analysis is complete.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from a healthcare provider (hcp) via manufacturer representative regarding a patient spinal therapy for c5/6 hernia.It was reported that after setting the left distractor, the welded part was broken and came off in about 2 clicks.Right distractor was used for dealing with it and no broken fragments left inside the patient.There was patient involved in the event and no further complications or symptoms were reported.On (b)(6) 2021 (rep): additionalinformation was received via manufacturer representative that there were no broken fragments left in side the patient.The levels implanted was c5/6 and there was no health damage in the patient.
 
Additional Manufacturer Narrative
Event there is no patient involved in the event.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Additional information was received via manufacturer representative that the part through which the pin passes and the main body part were separated.There is no patient involved in the event and product did not come in contact with the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIMLINE
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
MDR Report Key11955870
Report Number1030489-2021-00755
Device Sequence Number4403443
Product Code LXH
UDI-Device Identifier00681490520133
UDI-Public00681490520133
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number8756481
Device Catalogue Number8756481
Device Lot NumberPM14C002
Was Device Available for Evaluation? No
Date Returned to Manufacturer05/26/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/17/2021
Supplement Date Received by Manufacturer05/17/2021
Initial Report FDA Received Date06/08/2021
Supplement Report FDA Received Date06/16/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-