| Model Number |
TB-0510IC |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/16/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The subject device was not returned to olympus medical systems corp.(omsc) for evaluation.Therefore, the exact cause of the reported event could not be conclusively determined.A supplemental report will be submitted, if additional or significant information becomes available at a later time.
|
| |
|
Event or Problem Description
|
|
Olympus medical systems corp.(omsc)was informed by the customer that during the procedure using the subject device, the following event occurred.The probe was broken off.The customer presumed that it was in contact with the spreading clamps in use on the thyroid gland.The intended procedure was completed with another device.There was no patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This is a supplemental report to provide additional information.The subject device was returned to olympus medical systems corp.(omsc) for evaluation.As a result of checking the device the probe was broken at 17 mm from the distal end.Scratches were discovered in the probe.The crack was widening from the scratched area.There were no scratches or contact marks at the non-insulated area of the grasping section.The manufacturing record was reviewed and found no irregularities.Based on the past similar cases, omsc presumes that the event occurred due to the following occurrence mechanism.The probe came into contact with a metal object or surgical instruments while the output was activating in seal & cut mode.The probe received an output load in seal & cut mode or received a load when grasping tissue.As a result, the probe cracked from the scratch.The probe broke when added some load.The above device handling has warned in the instruction manual.
|
| |
|
Search Alerts/Recalls
|