| Model Number |
15 |
| Medical Device Problem Code |
Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/11/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Physio-control evaluated the customer's device and was unable to duplicate the reported issue.Physio replaced the device's battery pins and main keypad assembly and tightened the connections to the power pcb assembly as a precaution.After observing proper device operation through functional and performance testing, the device was returned to the customer for use.The cause of the reported issue could not be determined.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device is unexpectedly powering itself off and on.As a result, defibrillation therapy may be delayed or unavailable, if needed.There was no report of patient use associated with the reported event.
|
| |
|