• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH SAGITTAL SAW ATTACHMENT; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH SAGITTAL SAW ATTACHMENT; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT Back to Search Results
Model Number 532.021
Device Problems Mechanical Jam (2983); Noise, Audible (3273); Excessive Heating (4030)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/24/2021
Event Type  malfunction  
Manufacturer Narrative
Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.The actual device was returned for evaluation.During evaluation it was determined that the initial reported condition of the device was making a grinding noise and heating up was not confirmed.Therefore, the assignable root cause was not determined.However, during evaluation, it was confirmed that the device jammed/seized.It was further determined that the moving parts of the device did not move smoothly, there was corrosion/rusting/pitting, and foreign substance/debris.The assignable root cause was determined to be due to improper maintenance.(b)(4).
 
Event Description
It was reported that during that during an unspecified veterinary surgical procedure, it was discovered that the sagittal saw attachment device made a loud noise and then became really hot and seized up.It was reported that there was a thirty-minute delay to the surgical procedure and the surgeon had to remove the bone manually.There was no human patient involvement as the device was used in a veterinary procedure.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAGITTAL SAW ATTACHMENT
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key11958046
MDR Text Key280539323
Report Number8030965-2021-04420
Device Sequence Number1
Product Code HWE
UDI-Device Identifier10886982240583
UDI-Public10886982240583
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial
Report Date 05/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/08/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number532.021
Device Catalogue Number532.021
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/02/2021
Date Manufacturer Received05/24/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-