B5: additional information received; as per the physician the cause of the event is unknown.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Device evaluation summary: there was no damage observed to the applier catheter.An endoanchor was loaded in the applier housing.Blood residue was visible in the housing.The applier was advanced and exited the guide tip with no issue noted.The handle was opened, and no fluid or blood was observed within the handle.No corrosion was observed to the circuit board or connector pins.All wires and connectors within the handle appeared to be intact and properly connected.The battery was removed and measured 7.69v.A new battery was attached, and the unit powered up with the blue error flashing, the forward light unresponsive and the back light flashing.The eprom was read and presented the following codes (locn x code): oax05 battery low detection, 0bx0e apply ready, 0cx07 drive motor operation time-out, 00x17 fatal.The device was restarted in factory mode with the with the blue error on, the forward light flashing and the back light on.On pressing the forward button, the tip of the applier rotated, and it was not possible to deploy the endoanchor.The shaft was cut immediately proximal to the tip; blood was visible inside the shaft.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|