| Model Number |
773656 |
| Medical Device Problem Code |
Air/Gas in Device (4062)
|
| Health Effect - Clinical Codes |
Abdominal Distention (2601); Insufficient Information (4580)
|
| Date of Event |
06/04/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The incident sample has been requested but to date has not been received for evaluation. if the sample is received, or if additional information pertinent to the incident is obtained a follow-up report will be submitted. as part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.
|
| |
|
Event or Problem Description
|
|
Customer reports: the kangaroo bag that air is continuously getting into the line when running on the pump.
|
| |
|
Additional Manufacturer Narrative
|
|
H3 evaluation summary: the device history record (dhr) was reviewed and shows evidence that the product was released according to all established procedures and quality documentation.One used sample was received at the manufacturing site for evaluation.Visual and functional inspection was performed, and no failure was found.No root cause could be found related to the manufacturing/production process.The root cause could not be determined as the device functioned as expected.No formal investigation action is being taken since the failure mode could not be confirmed related to the manufacturing/production process.This complaint will be used for tracking and trending purposes.
|
| |
|