| Model Number |
0250070443 |
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/10/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device manufacture date is not known at this time.However, should it become available it will be provided in future reports.Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that the heat shrink and sheath melted.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: yellow heat shrink and sheath are melting, causing compromised insulation.No adverse consequences to patient.The failure(s) identified in the investigation is consistent with the complaint record.The probable root causes could be conductive irrigating or aspirating fluid used was inside the distal end, and contacted the probe and the sheath causing the sheath to melt; power set too high.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.Manufacture date is not known.
|
| |
|
Event or Problem Description
|
|
It was reported that the heat shrink and sheath melted.
|
| |
|
Search Alerts/Recalls
|