MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 SIG HP REV 4IN1 CUT BLK SZ4; KNEE INSTRUMENT : RESECTION BLOCKS/DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 SIG HP REV 4IN1 CUT BLK SZ4; KNEE INSTRUMENT : RESECTION BLOCKS/DEVICES Back to Search Results
Model Number 2011-03-003
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/26/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint #:(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that while trying to remove the pin out of the size 4 tc3 femoral cutting block, the pin got stuck and caused damage to the inside hole of the cutting block.Case was finished okay and without any problems or delays.Just need block replaced.No surgical delay.
 
Additional Manufacturer Narrative
Product complaint # : pc-(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: examination of the returned device could not confirm the pin was stuck inside the cutting block.However, the investigation revealed the inside of one of the cut block pin fixation holes was damaged and would not allow a pin to pass through.The investigation did not establish that a broader investigation or corrective action was necessary.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the product and lot code was provided, a manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIG HP REV 4IN1 CUT BLK SZ4
Common Device Name
KNEE INSTRUMENT : RESECTION BLOCKS/DEVICES
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key11958828
Report Number1818910-2021-12086
Device Sequence Number3754820
Product Code HTZ
UDI-Device Identifier10603295081241
UDI-Public10603295081241
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2011-03-003
Device Catalogue Number201103003
Device Lot NumberSO2021981
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/07/2021
Initial Date Received by Manufacturer 05/26/2021
Supplement Date Received by Manufacturer07/07/2021
07/15/2021
Initial Report FDA Received Date06/08/2021
Supplement Report FDA Received Date07/13/2021
07/16/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-