| Model Number |
8780 |
| Medical Device Problem Codes |
Difficult to Advance (2920); Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/04/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Information was received by a healthcare provider (hcp) via company representative (rep) regarding a patient receiving unknown drug via implantable infusion pump.It was reported that during the intrathecal baclofen (itb) implant, the physician used the needle to get intrathecal access.When the physician advanced the catheter, the catheter looped and traveled caudal rather than cephalad.When the physician tried to pull the catheter back it got stuck on the needle.The factors that may have led or contributed to the issue was unknown.A new catheter and needle was opened and she entered the intrathecal space more medially and was able to advance the new catheter without complications.The issue was resolve at the time of report.The patient's weight was asked and will not be made available.The patient's medical history includes multiple sclerosis.The patient's status at the time of report was alive - no injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The catheter was returned and analysis identified an anomaly to the catheter/guide wire.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|