• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 GLBL UNITE ANT BODY 135 SZ 12; GLOBAL UNITE IMPLANTS : SHOULDER HUMERAL STEM ACCESSORY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 GLBL UNITE ANT BODY 135 SZ 12; GLOBAL UNITE IMPLANTS : SHOULDER HUMERAL STEM ACCESSORY Back to Search Results
Model Number 1100-40-000
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 05/24/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint #:(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
The patient was revised to address infection and rotator cuff deficiency.No surgical delay reported.Doi: (b)(6) 2020.Dor: (b)(6) 2021.Left shoulder.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLBL UNITE ANT BODY 135 SZ 12
Common Device Name
GLOBAL UNITE IMPLANTS : SHOULDER HUMERAL STEM ACCESSORY
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key11960002
Report Number1818910-2021-12144
Device Sequence Number3565892
Product Code HSD
UDI-Device Identifier10603295004417
UDI-Public10603295004417
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K170748
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1100-40-000
Device Catalogue Number110040000
Device Lot Number9309460
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/24/2021
Supplement Date Received by Manufacturer07/06/2021
Initial Report FDA Received Date06/08/2021
Supplement Report FDA Received Date07/07/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GLBL UNITE ANT BODY 135 SZ 12; GLOBAL APG 48MM; GLOBAL UNITE HEAD 48X15 STD; GLBL UNITE ANT BODY 135 SZ 12; GLOBAL APG 48MM; GLOBAL UNITE HEAD 48X15 STD
Outcome Attributed to Adverse Event Required Intervention;
Patient Age38 YR
-
-