| Model Number |
977A275 |
| Medical Device Problem Codes |
Break (1069); Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/11/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Analysis of the lead found that the stylet coil was stretched due to a procedural non-conformance.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The health care professional reported via a manufacturer representative that the lead was attempted to be implanted.However, when the stylets were attempted to be changed out, neither (straight or curved) would thread all the way back to the lead tip.Both stylets were stuck at the 5th contact.This was during the procedure, so a different lead was used and two leads were ultimately placed. lead was removed from the body to attempt to thread the stylets.Neither stylet would thread outside of the body either.
|
| |
|