• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IMPRESS®; CATHETER, INTRAVASCULAR, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IMPRESS®; CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Model Number 00884450009185
Medical Device Problem Codes Fracture (1260); Material Separation (1562)
Health Effect - Clinical Codes Coma (2417); Foreign Body In Patient (2687)
Date of Event 03/24/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The suspect device is not expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
 
Event or Problem Description
The account alleges that the impress catheter tip broke off during a neurology vascular intervention.The tip of the catheter migrated along the vessel toward the brain.The patient was unconscious and comatose.An additional device was used to complete the procedure.
 
Additional Manufacturer Narrative
The suspect device was not returned for evaluation.The complaint could not be confirmed.The root cause could not be determined.The device history record was reviewed, and no exception documents were found.A search of the complaint database was performed and no similar complaints for this lot number were identified.Should the device be returned at a later date, the investigation will be re-opened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPRESS®
Common Device Name
CATHETER, INTRAVASCULAR, DIAGNOSTIC
MDR Report Key11962101
Report Number3010665433-2021-00038
Device Sequence Number1376903
Product Code DQO
UDI-Device Identifier00884450009185
UDI-Public00884450009185
Combination Product (Y/N)N
Initial Reporter CountryVM
PMA/510(K) Number
K053171
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2022
Device Model Number00884450009185
Device Catalogue Number510035B2
Device Lot NumberE1578092
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/29/2021
Supplement Date Received by Manufacturer08/02/2021
Initial Report FDA Received Date06/08/2021
Supplement Report FDA Received Date08/03/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/15/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-