Catalog Number 115307
Device Problems Break (1069); Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported during priming with a prismaflex m150 set, fluid leakage was observed.It was reported it was due to return tube damage.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Correction: b4/f8: date of this report in follow-up mdr #1 is being corrected from blank to 07/12/2021.
 
Manufacturer Narrative
Additional information: d9, h3, h6 and h10.The device was received for evaluation.This sample was visually inspected, and a leak test performed.A leak was observed from the effluent port due to a crack.The reported condition is verified.The cause of the crack was due to a transportation issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX M150
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key11962331
MDR Text Key259588421
Report Number8010182-2021-00188
Device Sequence Number1
Product Code KDI
Combination Product (y/n)Y
PMA/PMN Number
NI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 07/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number115307
Device Lot Number20K0904
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/14/2021
Initial Date Manufacturer Received 05/18/2021
Initial Date FDA Received06/08/2021
Supplement Dates Manufacturer Received07/12/2021
07/12/2021
Supplement Dates FDA Received07/12/2021
07/12/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-