MAUDE Adverse Event Report: ARTHROCARE CORP. DISP FIRSTPASS STR PASSR SELF; PASSER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHROCARE CORP. DISP FIRSTPASS STR PASSR SELF; PASSER Back to Search Results
Catalog Number 22-4038
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/12/2021
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event Description
It was reported that during use in an acl procedure, the jaw of the firstpass came loose from the instrument when the device was inside patient.The parts could be recovered.Procedure finished with s&n backup device.No surgical delay or complications to patient reported due this event.
 
Event Description
It was reported that during use in an acl procedure, the jaw of the firstpass came loose from the instrument.The broken piece was retrieved and the procedure was finished with a s&n backup device.No surgical delay or patient complications were reported.
 
Manufacturer Narrative
The reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review for the reported batch number concluded this was an isolated event.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of risk management files found that the reported failure was documented appropriately.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.B5: updated event description.H6: updated codes.
 
Manufacturer Narrative
The reported device, used in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection of the returned instrument shows no manufacturing abnormalities.The suture capture is detached and not returned.Product was out of the original packaging.No packaging returned.A functional evaluation revealed the needle will deploy when trigger is initiated.No suture capture to test sutures.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of risk management files found that the reported failure was documented appropriately.The complaint was confirmed.Factors that could have contributed to the reported event include: (1) excessive force (2) tissue thickness (3) damage or debris on the device tip between passes.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISP FIRSTPASS STR PASSR SELF
Type of Device
PASSER
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
Manufacturer (Section G)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key11963747
MDR Text Key255104224
Report Number3006524618-2021-00700
Device Sequence Number1
Product Code HWQ
UDI-Device Identifier00885556724545
UDI-Public885556724545
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 11/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/09/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/01/2023
Device Catalogue Number22-4038
Device Lot Number2065188
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received11/01/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/26/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age23 YR
Patient SexFemale
-
-