| Model Number |
ITHA54 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
05/26/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zimmerbiomet complaint number (b)(4).Patient identifier unknown / not provided.Age and date of birth unknown / not provided.Patient sex unknown / not provided.Weight unknown / not provided.Last/given name unknown / not provided.Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that while attempting to place a healing abutment in the patient's mouth, the screw portion fractured.The doctor was able to remove it without any damage to the implant.They were able to complete the procedure with another healing abutment.
|
| |
|
Event or Problem Description
|
|
No additional event information at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted to relay additional information and device evaluation.The following sections are being reported: h6: tyoe of investigation codes were added: 4109, 4110, 4111 and 3331.H6: investigation findings code was added: 3252.H6: investigation conclusions code was added: 4307.One certain® ep® healing abutment 4.1mm(d) x 5mm(p) x 4mm(h) (itha54) was returned for investigation.Visual evaluation of the as returned product identified that the device was fractured at the threaded region (screw portion).Functional testing could not be performed since the device was fractured.Per the applicable ifu, it is stated that improper technique can lead to device failure.Additionally, breakage may occur when device is loaded beyond its functional capability.Dhr review for the lot (1241150) had revealed no deviations nor non-conformances which could have caused or contributed to the reported event.All products were conforming at the time they left zimmer biomet.Lot was inspected and passed all acceptance criteria by qa.Complaint history review was performed for the reported lot number (1241150) for similar events (complaint category keywords: fracture: screw) and no other complaint was identified.August post market trending was reviewed and there were no actionable events or corrective actions for the reported event or product.Therefore, based on the available information, device malfunction did occur and the reported event was confirmed.
|
| |
|