Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the ascope had the lens tip broken.No case reported; therefore, there was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the reported device, intended for use in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection found distal tip damage, a scratched negative lens, and a cracked negative lens.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review concluded this was a repeat issue.The complaint was confirmed.Factors that could have contributed to the reported event include an impact event inconsistent with normal use.Internal complaint reference: (b)(4).
|
| |
|