• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ACE SURGICAL SUPPLY CO. INFINITY IHEX IMPLANT; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ACE SURGICAL SUPPLY CO. INFINITY IHEX IMPLANT; DENTAL IMPLANT Back to Search Results
Model Number 214710
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 05/05/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reported event has been determined to be a post-placement/pre-load failure.This implant loss suggests that the source of the problem was likely to stem from no osseointegration.Additionally, other factors that may have contributed to the implant failure includes bone condition, patient oral hygiene, overall health or behavior.Proper intended use of the implant is described in its instructions for use.
 
Event or Problem Description
Mobility was reported.Bone quality was identified as type ii and type iii.Time of loss in relation to the implantation was before prosthetic restoration.Primary stability was achieved.Osseointegration was achieved.Augmentation procedure was not performed at time of surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITY IHEX IMPLANT
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ACE SURGICAL SUPPLY CO.
1034 pearl street
brockton MA 02301
MDR Report Key11965894
Report Number1287163-2021-00872
Device Sequence Number14520454
Product Code DZE
UDI-Device Identifier00614950007189
UDI-Public00614950007189
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K103790
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 06/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device No Information
Device Model Number214710
Device Catalogue Number214710
Device Lot Number20120075
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/21/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/21/2021
Initial Report FDA Received Date06/09/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age89 YR
-
-