| Model Number |
S699GX |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/14/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during right index finger laceration repair, the needle was secured using needle driver and introduced to wound edge.Needle driver was able to loop the needle through wound up to the opposite side, however the base of the needle broke approximately 2mm from nylon suture.Another suture was opened to complete the case.There was no patient injury.
|
| |
|
Search Alerts/Recalls
|