• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS NORTH AMERICA LLC HEARTSTART MRX MONITOR/DEFIB; DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS NORTH AMERICA LLC HEARTSTART MRX MONITOR/DEFIB; DEFIBRILLATOR Back to Search Results
Model Number M3535A
Device Problem Sparking (2595)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/14/2021
Event Type  malfunction  
Event Description
It was reported that the device's paddles are sparking.There was reportedly no patient involvement.
 
Event Description
It was reported that the device's paddle are sparking.There was reportedly no patient involvement.
 
Manufacturer Narrative
The manufacturer's authorized field service engineer (fse) evaluated the device and confirmed the reported problem.Upon conclusion of the evaluation, it was determined that this was a malfunction of the paddle pocket plates, which was replaced, and the device was returned to full functionality.The device remains at the customer site and no further evaluation is warranted at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTSTART MRX MONITOR/DEFIB
Type of Device
DEFIBRILLATOR
Manufacturer (Section D)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer Contact
kara grubbs
22100 bothell everett highway
bothell, WA 98021
9095703538
MDR Report Key11966786
MDR Text Key255132827
Report Number3030677-2021-12077
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00884838000018
UDI-Public00884838000018
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
K031187
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM3535A
Device Catalogue NumberM3535A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date05/14/2021
Initial Date Manufacturer Received 05/14/2021
Initial Date FDA Received06/09/2021
Supplement Dates Manufacturer Received05/14/2021
Supplement Dates FDA Received07/12/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-