| Model Number |
203711 |
| Medical Device Problem Codes |
Failure to Osseointegrate (1863); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
04/22/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The reported event has been determined to be a post-placement/pre-load failure.This implant loss suggests that the source of the problem was likely to stem from no osseointegration.Additionally, other factors that may have contributed to the implant failure includes bone condition, patient oral hygiene, overall health or behavior.Proper intended use of the implant is described in its instructions for use.
|
| |
|
Event or Problem Description
|
|
Swelling and bleeding were reported.Bone quality was identified as type i.Time of loss in relation to the implantation was before prosthetic restoration.Primary stability was achieved.Osseointegration was not achieved.Augmentation procedure was not performed at time of surgery.
|
| |
|