The loaner device was previously returned to pentax medical from a customer on (b)(6)2021 and inspection of the unit was performed under service order (b)(4) where the quality control inspector documented the following codes on (b)(6) 2021: distal cap - fixed type failed epoxy seal integrity inspection, distal cap/ case cracked, passed dry leak test, passed wet leak test, middle lcb distal cover glass glue is missing, operation channel- primary mild resistance.The duodenoscope's repairs will include the distal case/cap, which will be replaced and/or resealed pursuant to the field correction, along with the following parts, and returned to the user upon completion: o-ring(0.5x1.3) imp-1.Model ed-3490tk, serial number (b)(4), has been routinely serviced at a pentax facility since the device was put into service on (b)(6) 2016.The endoscope is awaiting repair or approval by final qc as of (b)(6) 2021.
|