Model Number ED-3490TK
Medical Device Problem Code Loss of or Failure to Bond (1068)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Type of investigation: testing of actual/suspected device.Investigation findings: wear problem.Investigation conclusions: cause traced to user.
 
Event or Problem Description
The loaner device was previously returned to pentax medical from a customer on (b)(6)2021 and inspection of the unit was performed under service order (b)(4) where the quality control inspector documented the following codes on (b)(6) 2021: distal cap - fixed type failed epoxy seal integrity inspection, distal cap/ case cracked, passed dry leak test, passed wet leak test, middle lcb distal cover glass glue is missing, operation channel- primary mild resistance.The duodenoscope's repairs will include the distal case/cap, which will be replaced and/or resealed pursuant to the field correction, along with the following parts, and returned to the user upon completion: o-ring(0.5x1.3) imp-1.Model ed-3490tk, serial number (b)(4), has been routinely serviced at a pentax facility since the device was put into service on (b)(6) 2016.The endoscope is awaiting repair or approval by final qc as of (b)(6) 2021.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Common Device Name
VIDEO DUODENOSCOPE
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA  196-0012
MDR Report Key11967532
Report Number9610877-2021-00122
Device Sequence Number17220586
Product Code FDT
UDI-Device Identifier04961333229611
UDI-Public04961333229611
Combination Product (Y/N)N
Initial Reporter StateMT
Initial Reporter CountryUS
PMA/510(K) Number
K161222
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Remedial Action Recall,Repair,Inspection
Type of Report Initial
Report Date (Section B) 05/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model NumberED-3490TK
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/28/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/14/2021
Initial Report FDA Received Date06/09/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
NOT ISSUED YET
Patient Sequence Number1
-
-