As reported by user facility: the venous line detached from the catheter two (2) hours into treatment causing significant blood loss of 500+ ml.The venous pressure reading of the dialysis machine was around 200 mmhg during treatment and then dropped rapidly to 40 mmhg when the venous line detached.Since the lower alarm limit of the dialog+ machine was set to 20 mmhg, it did not trigger an alarm.The nurse responded within one (1) minute of the pressure drop to find that the venous line had become disconnected.The site was not covered.It was in plain sight.The patient received cpr and was transported to the hospital.The patient did pass away several days later.Please note that this is the same patient and incident as reported in mdr # 2521402-2021-00024.Additional importer narrative: the dialog+ dialysis machine was technically inspected by the customer's technician after the described event.It was confirmed that it operated as intended and has been cleared for use by the customer.Dependent on the pressure drop and the alarm limits set, the pressure monitoring system of a dialysis machine does not reliably detect a disconnection or dislodgement of the venous patient access.Therefore, the instructions for use of the dialog+ dialysis machine shows the following warning: risk to patient due to blood loss if cannulas get disconnected or dislodged! standard monitoring function of the dialysis machine cannot ensure detection if the cannulas get disconnected or dislodged.Ensure that the access to the patient always remains fully visible during therapy.Ensure that cannulas are adequately fixed.Regularly check patient access.Venous lower limit should be set to greater than or equal to 20 mmhg in tsm (technical service mode).
|