It was reported that, during thr surgery, the mechanism of a ratchet handle was broken while the instrument was handled outside the patient.No significant surgical delay was reported as a smith and nephew backup ratchet handle was available.Neither patient injury nor other complications were reported.
|
|
H6: the device, used in treatment, was not returned for evaluation.Therefore, the product analysis and the reported event could not be confirmed at this time.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
|