| Model Number |
IPN008820 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/19/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Qn#(b)(4).The device history review for the product visist at 35w non-sterile lot# 73g2000119 investigation did not show issues related to the complaint.The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
|
| |
|
Event or Problem Description
|
|
Stapler lost staples while being used.Fell from stapler.
|
| |
|
Event or Problem Description
|
|
Stapler lost staples, while being used fell from stapler.
|
| |
|
Additional Manufacturer Narrative
|
|
Qn# (b)(4).The customer returned one unit 528236, visistat 35w non-sterile for investigation.The returned stapler was visually examined with and without magnification.Visual examination of the returned sample revealed, that there was one loose staple remaining in the cover block.The pusher and shuttle were both loose within the cover block as well.Upon further inspection, it was observed, that the bottom and rail were detached on one side of the cover block.It appears that the bottom was not properly welded onto the cover block.Which allowed the components within the cover block to become loose and fall out of the stapler.A nonconformance has been opened by the manufacturing site, as a result of this issue.
|
| |
|