Model Number 1219-36-156
Medical Device Problem Codes Fracture (1260); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 05/24/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the patient had his replaced 14 years ago by another surgeon at another hospital which was not known at the time i asked during surgery.This patient did well until recently where he came in to see the surgeon.The x-rays indicated the femoral head was articulating on the cup.The surgeon and the patient decided that a revision would be done to correct the issue.During the surgery the 56mm acetabular sector cup, 56mm marathon liner, and a 36mm 8.5 hip ball were explanted.The aml stem, lott and cat # unknown was left in vitro.The acetabular liner was broken and the hip ball was rubbing against the acetabular cup which caused significant damage to the femoral head and liner.A new competitor cup was implanted.Doi: 14 years ago - dor: (b)(6) 2021 (left hip).
 
Additional Manufacturer Narrative
Product complaint (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN MAR +4 10D 36IDX56OD
Common Device Name
PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key11968604
Report Number1818910-2021-12253
Device Sequence Number3674095
Product Code LPH
UDI-Device Identifier10603295014591
UDI-Public10603295014591
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K983014
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1219-36-156
Device Catalogue Number121936156
Device Lot NumberDD9F81000
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/24/2021
Supplement Date Received by Manufacturer07/13/2021
Initial Report FDA Received Date06/09/2021
Supplement Report FDA Received Date07/14/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DELTA CER HEAD 12/14 36MM +8.5; PINN MAR +4 10D 36IDX56OD; PINNACLE SECTOR II CUP 56MM; UNKNOWN HIP FEMORAL STEM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age81 YR
-
-