MAUDE Adverse Event Report: MEDTRONIC HEART VALVES DIVISION SIMULUS SEMI-RIGID ANNULOPASTY RING; RING, ANNULOPLASTY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC HEART VALVES DIVISION SIMULUS SEMI-RIGID ANNULOPASTY RING; RING, ANNULOPLASTY Back to Search Results
Model Number 800SR32
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Code Mitral Valve Insufficiency/ Regurgitation (4451)
Date of Event 05/20/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product analysis: no product was returned.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information that during the implant of this 32mm mitral annuloplasty ring, it was explanted and replaced with an annuloplasty product of a different model due to mild regurgitation caused by a sizing mismatch.It was reported the patient had anatomical features that caused difficulties in sizing the valve.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
Conclusion: based on the reported information at this time, the root cause of the sizing mismatch was patient anatomy.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIMULUS SEMI-RIGID ANNULOPASTY RING
Common Device Name
RING, ANNULOPLASTY
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
MDR Report Key11968715
Report Number2025587-2021-01833
Device Sequence Number4543311
Product Code KRH
UDI-Device Identifier00643169182271
UDI-Public00643169182271
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K072655
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/16/2025
Device Model Number800SR32
Device Catalogue Number800SR32
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/21/2021
Supplement Date Received by Manufacturer08/02/2021
Initial Report FDA Received Date06/09/2021
Supplement Report FDA Received Date08/03/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
Patient Weight75
-
-