| Model Number |
977D260 |
| Medical Device Problem Code |
Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/30/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The returned device was subjected to a series of standard tests that include but is not limited to visual inspection and electrical testing.Analysis determined that foreign material was blocking the insertion of the stylet.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The healthcare provider (hcp) reported via the manufacturer representative that while changing out stylets they were unable to feed stylet into the lead.The hcp said he tried multiple stylets and they would not go all the way in.No external or patient factors were noted to have contributed to the issue.The hcp replaced the lead and the issue was resolved.There were no symptoms as no patient was involved.
|
| |
|