Model Number 977D260
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/30/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The returned device was subjected to a series of standard tests that include but is not limited to visual inspection and electrical testing.Analysis determined that foreign material was blocking the insertion of the stylet.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The healthcare provider (hcp) reported via the manufacturer representative that while changing out stylets they were unable to feed stylet into the lead.The hcp said he tried multiple stylets and they would not go all the way in.No external or patient factors were noted to have contributed to the issue.The hcp replaced the lead and the issue was resolved.There were no symptoms as no patient was involved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURESCAN
Common Device Name
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
MDR Report Key11970287
Report Number2649622-2021-11525
Device Sequence Number17235131
Product Code LGW
UDI-Device Identifier00763000006570
UDI-Public00763000006570
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/11/2024
Device Model Number977D260
Device Catalogue Number977D260
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/19/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/08/2021
Initial Report FDA Received Date06/09/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-