| Model Number |
FRED4528-PMA |
| Medical Device Problem Codes |
Separation Failure (2547); Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/10/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A search for non-conformances associated with the reported part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The stent was returned without the original pusher.However, the stent was able to fully open and detach from a new pusher wire during the investigation.The damage to the tantalum wire is consistent with the stent being retrieved with a snare.The non-returned pusher is a critical component for this investigation; without the return and physical evaluation of the pusher, the complaint is considered non-verifiable.
|
| |
|
Event or Problem Description
|
|
It was reported that a flow diverter device (fred 27 stent) used for treatment of a paraclinoid aneurysm would not release from the delivery pusher wire.The flared ends of the stent were open just enough to be loose from the pusher wire but the stent would not open fully.A snare was used to removed the device without any issue and the case was ended without treatment of the aneurysm.The patient suffered no consequences due to the procedure or post procedure and was discharged the next day.
|
| |
|
Search Alerts/Recalls
|