• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO MANUFACTURING NETHERLANDS TECNIS IOL; EXTENDED DEPTH OF FOCUS INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO MANUFACTURING NETHERLANDS TECNIS IOL; EXTENDED DEPTH OF FOCUS INTRAOCULAR LENS Back to Search Results
Model Number ZXT375
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Visual Disturbances (2140); Halo (2227)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event: unknown, not provided.If explanted, give date: not applicable, as lens remains implanted.(b)(4).The device is not returning for evaluation as lens remains implanted; therefore, a failure analysis of the complaint device cannot be completed.A review of the device/lot history record and complaint trending for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.Attempts have been made to obtain missing information; however, no definitive response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was initially reported that a doctor is possibly going to explant an intraocular lens (iol) that was implanted in 2017 due to the patient experiencing glare since surgery.Upon follow-up it was confirmed that the patient has a lens implanted in both the left and right eye.It was noted that the patient is also experiencing halos, but it is unknown when this symptom began.The patient's visual concerns are reported to be debilitating and are significantly interfering with daily activities.Because of this, the patient wants the lenses removed.No other information was provided.This mdr is for report on suspect product in left eye.A separate report will be submitted for the suspect product in right eye.
 
Additional Manufacturer Narrative
Additional information: upon further follow-up it was noted that the suspect zxt375 +12.5 serial number (sn) (b)(6) was explanted and replaced with zcu375 +13.0 sn (b)(6).The tecnis inserter with platinum cartridge was used.The wound was slightly enlarged but no suture was required.No wound leak.Patient is doing well.Appropriate section in the mdr updated accordingly.Section d6: explant date: (b)(6), 2021.Device evaluation: product evaluation was not performed because the product has not been returned.Complaint report cannot be verified.Manufacturing record review: the manufacturing process record was evaluated and revealed that the product was manufactured and released according to specifications.A search revealed that one complaint was found as part of this production order although this complaint has been voided.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Device evaluation: visual inspection under magnification revealed two parts of an iol (one part missing).Viscoelastic residue was found on the optic body and haptic, which is consistent with a lens that was handled during explant.Based on the return condition of the lens, no further product evaluation could be performed.The complaint issue could not be confirmed, and no product deficiency could be identified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Corrected data: upon further review it was discovered that the section h4: device manufacture date was incorrectly populated.Correct date is july 19, 2017.Section below updated in this supplemental mdr submission.Section h4: device manufacture date: july 19, 2017.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
EXTENDED DEPTH OF FOCUS INTRAOCULAR LENS
Manufacturer (Section D)
AMO MANUFACTURING NETHERLANDS
van swietenlaan 5
groningen, groningen 9728 NX
NL  9728 NX
MDR Report Key11972639
Report Number9614546-2021-07188
Device Sequence Number4492106
Product Code POE
UDI-Device Identifier05050474606821
UDI-Public(01)05050474606821(17)220719
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 10/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/19/2022
Device Model NumberZXT375
Device Catalogue NumberZXT375U125
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/27/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 05/12/2021
Supplement Date Received by Manufacturer07/22/2021
10/19/2021
10/19/2021
Initial Report FDA Received Date06/10/2021
Supplement Report FDA Received Date08/03/2021
10/19/2021
10/19/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-