MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC FS LITE-NO CODING; BLOOD GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC FS LITE-NO CODING; BLOOD GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 71026-70
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/26/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product has been requested back for investigation.A follow-up report will be submitted once additional information is obtained.All pertinent information available to abbott diabetes care has been submitted.
 
Event or Problem Description
A customer reported receiving erroneous glucose results from an abbott diabetes care device.The results when plotted on a parkes error grid fell into either the c, d, or e zone.There was no report of death, serious injury, or mistreatment associated with this event.
 
Additional Manufacturer Narrative
No product has been returned.Extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.Dhrs for the freestyle freedom lite meter were reviewed and the dhr showed the freestyle freedom lite meter passed all tests prior to release.Dhrs for the freestyle strips were reviewed and the dhr showed the freestyle strips passed all tests prior to release.Retain testing was performed for freestyle strips and all units performed in specification and passed.If the product is returned, a physical investigation will be performed, and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
 
Event or Problem Description
A customer reported receiving erroneous glucose results from an abbott diabetes care device.The results when plotted on a parkes error grid fell into either the c, d, or e zone.There was no report of death, serious injury, or mistreatment associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FS LITE-NO CODING
Common Device Name
BLOOD GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
MDR Report Key11973022
Report Number2954323-2021-71872
Device Sequence Number4256865
Product Code NBW
UDI-Device Identifier00699073710260
UDI-Public00699073710260
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K092602
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date07/31/2022
Device Model Number71026-70
Device Catalogue Number71026
Device Lot Number1130564
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/26/2021
Supplement Date Received by Manufacturer06/10/2021
Initial Report FDA Received Date06/10/2021
Supplement Report FDA Received Date06/11/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
Patient Age82 YR
Patient Weight93
-
-