DEPUY SYNTHES PRODUCTS LLC 12.4 CM HEAVY DUTY LONG ATTACHMENT; MOTOR, DRILL, ELECTRIC - ATTACHMENT
|
Back to Search Results |
|
| Model Number |
LONG-HD-G1 |
| Medical Device Problem Code |
Excessive Cooling (2932)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, it was determined that the reported condition was confirmed.The assignable root cause was determined to be due to component failure from normal wear.Udi: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported from (b)(6) that during service and evaluation, it was determined that the attachment device had vibration and heat.It was further determined that the device failed pretest for vibration and temperature assessment.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2021.All available information has been disclosed.If additional information should become available, an additional report will be submitted accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|