• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH T-PAL SPACER APPLICATOR HANDLE; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH T-PAL SPACER APPLICATOR HANDLE; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR Back to Search Results
Catalog Number 03.812.001
Medical Device Problem Code Compatibility Problem (2960)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/02/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2021, at the distal tip there were markings that made it difficult to insert the inner shaft and trial implant.The alignment guide had a defect in the distal part and the cement cannula would not go through it.The issue was identified during pre-op.Another product was used to complete the surgery.There was no reported surgical delay.This report involves one (1) t-pal spacer applicator handle.This is report 2 of 3 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H3, h4, h6: part # 03.812.001; lot # l042558; release to warehouse date: november 10, 2016; manufacturer: hagendorf; supplier: (b)(4).A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Visual inspection: applicat out shaft (part# 03.812.001, lot# l042558, qty# 1) was returned at us customer quality (cq).Upon visual inspection it was observed that received outer shaft was received assembled with inner shaft.Functional test: functional testing of the received device was performed at cq.It was observed that applicator inner shaft was stuck inside the outer shaft and was unable to disassemble as intended to.Can the complaint be replicated with the returned device(s)? yes dimensional inspection: a dimensional inspection was not performed due to post-manufacturing damage document/specification review: the following document(s) was reviewed: applicator handle cpl: current and manufactured revisions applicator inner shaft cpl: current revision no design issues or discrepancies were found during this investigation.Complaint confirmed? yes.Investigation conclusion: the complaint can be confirmed for applicat out shaft.A definitive root cause could not be identified for the reported problem.There was no indication that a design or manufacturing issue contributed to the complaint.No design issues were observed during the document/specification review.Based upon these results, no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T-PAL SPACER APPLICATOR HANDLE
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key11975732
Report Number8030965-2021-04772
Device Sequence Number13741471
Product Code MAX
UDI-Device Identifier07611819405813
UDI-Public(01)07611819405813
Combination Product (Y/N)N
Initial Reporter CountryNO
PMA/510(K) Number
K151276
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.812.001
Device Lot NumberL042558
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/01/2021
Initial Date Received by Manufacturer 06/03/2021
Supplement Date Received by Manufacturer09/01/2021
Initial Report FDA Received Date06/10/2021
Supplement Report FDA Received Date09/01/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
APPLICAT OUT SHAFT.; APPLICAT OUT SHAFT.; APPLICAT OUT SHAFT.; APPLICATOR INNER SHAFT.; APPLICATOR INNER SHAFT.; UNK - INSERTION INSTRUMENTS.; APPLICAT OUT SHAFT; UNK - INSERTION INSTRUMENTS
-
-