• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION CAPTIFLEX; SNARE, FLEXIBLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION CAPTIFLEX; SNARE, FLEXIBLE Back to Search Results
Model Number M00562402
Medical Device Problem Codes Break (1069); Entrapment of Device (1212); Failure to Cut (2587)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/01/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The exact date of the event is unknown.The provided event date of (b)(6) 2021 was chosen as a best estimate based on the date that the manufacturer became aware of the event.The complainant was unable to provide the device lot number.Therefore, the manufacture and expiration dates are unknown.(b)(4).The complainant indicated that the device was disposed and will not be returned for evaluation; therefore, a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that a captiflex medium oval flexible snare was to be used during an upper endoscopy procedure performed on an unknown date.It was reported that during the procedure, the handle broke while they were trying to close the snare with cautery.The snare was halfway through the polyp and was stuck on the polyp.The sheath was removed and another instrument was used to try to get the snare out from being embedded in the polyp.The patient was transported to yale main campus and during transport the snare came off.Reportedly, the procedure was unsuccessful.There were no patient complications reported as a result of this event.The patient condition following the procedure was reported to be fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPTIFLEX
Common Device Name
SNARE, FLEXIBLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key11979804
Report Number3005099803-2021-02646
Device Sequence Number3661055
Product Code FDI
UDI-Device Identifier08714729501640
UDI-Public08714729501640
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K131700
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 06/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM00562402
Device Catalogue Number6240-40
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/14/2021
Initial Report FDA Received Date06/10/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-