OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II VIDEO SYSTEM CENTER Back to Search Results
Model Number CV-180
Device Problem Erratic or Intermittent Display (1182)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/07/2021
Event Type  malfunction  
Event Description
Olympus medical systems corp.(omsc) was informed from the user that the image of the subject device was flickered when the camera head connected to the subject device was slightly moved during preparation for use.The user took the precaution on it during the procedure and completed the procedure with using the same set of equipment, and without the patient injury.
 
Manufacturer Narrative
The subject device is planned to be returned to olympus but has not been returned to olympus yet.Therefore, olympus could not investigate the subject device.The exact cause of the reported event could not be conclusively determined at this time.If additional information is received, this report will be supplemented.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.The subject device has not been returned to omsc but was returned to olympus malaysia for evaluation.Olympus malaysia checked the subject device and duplicated the reported phenomenon.It was found the video connector socket failure of the subject device which made image flicking.Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.Based upon the information from olympus malaysia, omsc surmised that the reported phenomenon might have occurred due to loosen connector with the abrasion in the lock and contact pin of the video connector socket of the subject device because it has been more than 10 years since the subject device manufacturing.If additional information becomes available, this report will be supplemented.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II VIDEO SYSTEM CENTER
Type of Device
VIDEO SYSTEM CENTER
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key11982556
MDR Text Key280683072
Report Number8010047-2021-07396
Device Sequence Number1
Product Code FAJ
Combination Product (y/n)N
PMA/PMN Number
K133538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup
Report Date 07/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCV-180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/18/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/18/2021
Initial Date FDA Received06/11/2021
Supplement Dates Manufacturer Received06/24/2021
Supplement Dates FDA Received07/12/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-