MAUDE Adverse Event Report: MEDTRONIC XOMED INC. XPS HANDPIECE - VISAO; DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) I

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC XOMED INC. XPS HANDPIECE - VISAO; DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) I Back to Search Results
Model Number 3334800T
Medical Device Problem Codes Fluid/Blood Leak (1250); Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/18/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As a result of the inspection, handpiece was found overheating at 1 minute.Also, the handpiece generates abnormal noise and the inside has become discolored to black.But the reported oil leaking couldn't be duplicated.Pre-temp results at 1 minute: rear 29.4 c, middle 34.9 c and nose 61.8 c.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A health care provider reported that the handpiece was overheating and the oil was leaking during tympanoplasty.There was no patient or staff impact.
 
Additional Manufacturer Narrative
H3: additional analysis: an internal inspection confirmed that parts such as the middle housing, shaft, stainless steel balls, collar, o-ring, nose, and bearings had deteriorated due to dirt and rust.The device has been disassembled and cleaned.Parts such as motor cable assembly, middle housing, shaft, stainless steel balls, collar, o-ring, nose, and bearings have been replaced.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XPS HANDPIECE - VISAO
Common Device Name
DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) I
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
MDR Report Key11983035
Report Number1045254-2021-00292
Device Sequence Number16918797
Product Code ERL
UDI-Device Identifier00885074107349
UDI-Public00885074107349
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K041523
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3334800T
Device Catalogue Number3334800T
Device Lot Number37091800
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/18/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/18/2021
Supplement Date Received by Manufacturer07/27/2021
Initial Report FDA Received Date06/11/2021
Supplement Report FDA Received Date08/12/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-