Model Number 3660 |
Device Problem
Unexpected Shutdown (4019)
|
Patient Problem
Inadequate Pain Relief (2388)
|
Event Date 12/21/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
Date of event is estimated.
|
|
Event Description
|
It was reported the patient experienced unexpected shutdown of stimulation with their ipg.Troubleshooting was unable to resolve the issue.In turn, surgical intervention may be pending to address the issue.
|
|
Event Description
|
Additional information received indicated the patient underwent surgical intervention wherein the ipg was explanted and replaced to address the issue.
|
|
Manufacturer Narrative
|
The reported observation of ¿therapy turning off unintentionally¿ was not confirmed or duplicated during analysis.The cause of the reported observation was not ascertained; the device exhibited normal device characteristics.The implantable pulse generator (ipg) diagnostic logs do show during the implant period stimulation was turned off and on repeatedly.It was also confirmed that once the device magnet mode was programmed to ¿ignore¿ while still implanted no further magnet events were logged, which is normal operation.The device was subjected to a functional test on automated test equipment (ate).This tester verifies the electrical performance of the control/communication circuitry, and output signal integrity.The ate test specifically tests the ipg output on all 16 electrodes including the ipg can.The magnet function of the ipg is tested during the ate test.All portions of ate testing passed.
|
|
Search Alerts/Recalls
|