| Model Number |
540-34-150 |
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
05/14/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
|
| |
|
Event or Problem Description
|
|
Second revision surgery - the humeral component was loose.Surgeon removed the humeral component from djo and put in a zimmer biomet segmental revision system humeral stem.The new condyle kit was to link that new stem with our ulnar stem.
|
| |
|
Event or Problem Description
|
|
Revision surgery - the humeral component was loose.Surgeon removed the humeral component from djo and put in a zimmer biomet segmental revision system humeral stem.The new condyle kit was to link that new stem with our ulnar stem.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer narrative: the reason for this revision surgery was reported as loose stem.The previous surgery and the surgery detailed in this event occurred 10.7 months apart.Initial or prolonged hospitalization was required.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to loose stem.There were no findings during this evaluation that indicate the reported devices were defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may contribute to an event that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
|
| |
|
Search Alerts/Recalls
|