• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR ARMADA 14 PTA CATHETER; PERIPHERAL DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR ARMADA 14 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Model Number A2030-060
Medical Device Problem Codes Inflation Problem (1310); Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/19/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A visual and functional inspection was performed on the returned balloon catheter and the reported inflation issue was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents and/or complaints reported from this lot.The device was prepped prior to use without any leak or ruptures noted, which would suggest that the device was not damaged prior to use.The investigation determined the reported inflation issue, noted balloon rupture and balloon scratch appear to be related to operational circumstances of the procedure.Based on the reported information and returned analysis, it is likely that during advancement through the severe calcified and mild tortuosity anatomy the balloon outer surface became damaged resulting in the noted balloon rupture/leak during inflation.The noted scratch at the rupture site also suggests interaction causing damage to the outer surface of the balloon material.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event or Problem Description
It was reported that the procedure was to treat a lesion with severe calcification and mild tortuosity in the left anterior tibial artery.The 3.0 x 60 mm armada 14 balloon catheter crossed the lesion; however, during an attempt to inflate the balloon to nominal pressure, the pressure did not increase.Therefore, the balloon catheter was removed without issue.A new armada 14 balloon was used successfully.There were no adverse patient effects and no clinically significant delay in the procedure.The returned device analysis identified a pinhole in the balloon material 2.7cm distal to the proximal marker.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 14 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key11986313
Report Number2024168-2021-04933
Device Sequence Number17333493
Product Code DQY
UDI-Device Identifier08717648156083
UDI-Public08717648156083
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K102705
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2022
Device Model NumberA2030-060
Device Catalogue NumberA2030-060
Device Lot Number9071641
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/29/2021
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/20/2021
Initial Report FDA Received Date06/11/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-