• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN 5FR TRPL-LUMN UMB VSL CATH; CATHETER, UMBILICAL ARTERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN 5FR TRPL-LUMN UMB VSL CATH; CATHETER, UMBILICAL ARTERY Back to Search Results
Model Number 8888160648
Medical Device Problem Codes Crack (1135); Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/31/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The complainant indicated that the device will not be returned for evaluation; therefore, a failure analysis is not available, and we are not able to determine the relationship between this device and the cause for this event.As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.If additional information or the sample is received, the investigation will be reopened and responded to accordingly.
 
Event or Problem Description
The customer states that a cardiac patient had a triple lumen umbilical vessel catheter and after placement, multiple fluid boluses and drops were ordered.The providers requested the bedside nurse hand push boluses instead of running them on a pump.The nurses had a hard time pushing the meds because there was a lot of resistance.It was discovered that there was a crack in the tertiary lumen after the echo tech felt the fluid from the boluses splashing him.The neonatal nurse practitioner was notified, and replaced the triple lumen umbilical vessel catheter with another.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5FR TRPL-LUMN UMB VSL CATH
Common Device Name
CATHETER, UMBILICAL ARTERY
Manufacturer (Section D)
COVIDIEN
edificio b20 calle #2 zona fra
alajuela
CS 
MDR Report Key11986376
Report Number3009211636-2021-00753
Device Sequence Number4637114
Product Code FOS
UDI-Device Identifier20884527005199
UDI-Public20884527005199
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Risk Manager
Type of Report Initial
Report Date (Section B) 06/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number8888160648
Device Catalogue Number8888160648
Device Lot Number1921900151
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/03/2021
Initial Report FDA Received Date06/11/2021
Patient Sequence Number1
-
-