• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT GMBH MAGNESIUM; PHOTOMETRIC METHOD, MAGNESIUM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT GMBH MAGNESIUM; PHOTOMETRIC METHOD, MAGNESIUM Back to Search Results
Model Number 03P6822
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/15/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.Patient identifier complete entry = sample id (b)(6).All available patient information was included.Additional patient details are not available.
 
Event or Problem Description
The customer observed a falsely elevated magnesium result for one patient on an architect c8000 analyzer.The following data was provided (customer¿s reference range is 1.5-2.3 mg/dl): sample id (b)(6) initial result was 6.2, repeats were 1.8 and 1.8 mg/dl, when repeating on an architect c4000 the result was 1.7 mg/dl.There was no impact to patient management reported.
 
Additional Manufacturer Narrative
The complaint investigation for a falsely elevated architect magnesium result included a review of data and information provided by the customer, search for similar complaints, ticket trending review, labeling review, and device history record review.Trending review determined no related trend for the increased result for the product.Return testing was not completed as returns were not available.Data provided by the customer showed that the sample was retested on the complaint instrument, and on a different instrument, with normal results that were similar between the instruments.Manufacturing documentation was reviewed and did not identify any issues.Labeling was reviewed and sufficiently addresses the customer's issue.Based on the information provided and abbott diagnostics¿ complaint investigation, no systemic issue or deficiency for the architect magnesium, lot number 87180un21, was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGNESIUM
Common Device Name
PHOTOMETRIC METHOD, MAGNESIUM
Manufacturer (Section D)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM  65205
MDR Report Key11987237
Report Number3002809144-2021-00375
Device Sequence Number4636529
Product Code JGJ
UDI-Device Identifier00380740161620
UDI-Public00380740161620
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K181748
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/10/2022
Device Model Number03P6822
Device Catalogue Number03P68-22
Device Lot Number87180UN21
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/15/2021
Supplement Date Received by Manufacturer06/25/2021
Initial Report FDA Received Date06/11/2021
Supplement Report FDA Received Date07/01/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARC C8 PROC MOD, 01G06-11, C803137.; ARC C8 PROC MOD, 01G06-11, C803137.
-
-