SMITH & NEPHEW, INC. LENS INTEGRATED SYSTEM WIFI VERSION; LAPAROSCOPE, GENERAL & PLASTIC SURGERY
|
Back to Search Results |
|
Model Number 72204354 |
Device Problems
No Display/Image (1183); Unexpected Shutdown (4019)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Internal complaint reference: (b)(4).
|
|
Event Description
|
It was reported that, during setup, the light on the lens integrated system kept turning off.It is unknown if a backup device was available.No delay nor patient injury was reported.No other complications were reported.
|
|
Manufacturer Narrative
|
Internal complaint reference: (b)(4).The reported device, intended for use in treatment, was received for evaluation.There was no relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.The complaint of light output could not be confirmed.Lens ccu passed all functional testing and functions perform as expected.The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.No containment or corrective actions are recommended at this time.
|
|