An endurant ii aui stent graft system was implanted in the patient for the endovascular treatment of an endoleak type 2a caused by a talent aortic stent.Approximately 64 post procedure, the patient presented emergently to the hospital with back pain.Abdominal ultrasound was performed, where an endoleak type 1a was noted.The event was treated with the implantation of 10 aptus endoanchors placed into the endurant aui in the aortic neck.The event was resolved.As per the physician, cause of the event was patient's anatomy possibly due to aortic dilation.No additional clinical sequalae were reported and the patient is fine.
|
|
It was reported that the reason the endurant aui was implanted was likely a type ia endoleak and not a type ii endoleak as previously reported.This information has not been confirmed as the sales representative was not present at that case.It was confirmed the endurant aui was treating a type ia endoleak on the talent stent graft.It was reported the ia endoleak was cause by neck dilation and was resolved with the implant of the aui stent graft.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|