• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CORDIS CORPORATION STENT PER 30MM 4.0MM LARGE; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CORDIS CORPORATION STENT PER 30MM 4.0MM LARGE; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number P308
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/12/1999
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
According to the literature article by slonim, s.M., dake, m.D., razavi, m.K., kee, s.T., samuels, s.L., rhee, j.S., & semba, c.P.(1999).Management of misplaced or migrated endovascular stents.Journal of vascular and interventional radiology : jvir, 10(7), 851¿859.Https://doi.Org/10.1016/s1051-0443(99)70127-2, a palmaz p308 stent was intended to be placed in the left common iliac vein (l civ) but migrated to the inferior vena cava (ivc) due to rupture of an unknown balloon.The stent was removed percutaneously with another unknown 3-5mm balloon catheter.The product was not returned for analysis.No lot number was provided therefore a product history record (phr) review could not be generated.The reported ¿stent dislodged - in patient¿ and ¿balloon-sds burst - in patient¿ could not be confirmed as the device was not returned for analysis.The exact cause could not be determined.Vessel characteristics are unknown.As no lot number was supplied a phr could not be completed.According to the safety information in the instructions for use ¿the palmaz balloon-expandable stent for the iliac arteries is 316l stainless steel, slotted tube.The stent should be crimped over the recommended balloon delivery catheter.Measure the diameter of the lesion to determine the appropriate size stent and delivery system.¿ the recommended balloon was the maxi ds.It is not known if the recommended balloon was used in this case.The limited information available does not suggest a design or manufacturing related cause for the reported event; therefore, no corrective/preventive action will be taken at this time.
 
Event or Problem Description
According to the literature article by slonim, s.M., dake, m.D., razavi, m.K., kee, s.T., samuels, s.L., rhee, j.S., & semba, c.P.(1999).Management of misplaced or migrated endovascular stents.Journal of vascular and interventional radiology : jvir, 10(7), 851¿859.Https://doi.Org/10.1016/s1051-0443(99)70127-2, a palmaz p308 stent was intended to be placed in the left common iliac vein (l civ) but migrated to the inferior vena cava (ivc) due to rupture of an unknown balloon.The stent was removed percutaneously with another unknown 3-5mm balloon catheter.The device will not be returned for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STENT PER 30MM 4.0MM LARGE
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
CORDIS CORPORATION
14201 nw 60 avenue
miami lakes FL 33014
MDR Report Key11991631
Report Number9616099-2021-04634
Device Sequence Number16961152
Product Code FGE
UDI-Device Identifier20705032057292
UDI-Public20705032057292
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K911581
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source literature
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberP308
Device Catalogue NumberP308
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/25/2021
Initial Report FDA Received Date06/14/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN 3-5MM BALLOON CATHETER; UNKNOWN BALLOON CATHETER
Patient Age67 YR
-
-