WRIGHTS LANE SYNTHES USA PRODUCTS LLC 3.5MM LCP PLATE 12 HOLES 163MM; PLATE, FIXATION, BONE
|
Back to Search Results |
|
| Model Number |
223.621 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Nonunion/Delayed-union Bone Fracture (2369)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Without a valid lot number the device history records review could not be completed.Complainant device is not expected to be returned for manufacturer review/investigation.The investigation could not be completed; no conclusion could be drawn, as no product was received.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on an unknown date, the patient underwent revision of a midshaft nonunion in the left humerus and right tibia.In patient's humerus, dynamic compression plates with limited bone contact (lc-dcp) and narrow dynamic compression plates with limited bone contact (lc-dcp) were removed, along with twelve cortex screws.Patient was revised to a new lcdcp large fragment plate and bone graft harvested from the patient¿s tibia.In patient's tibia, locking compression plate (lcp) and locking compression plate (lcp) tubular plate were removed, along with twenty-one cortex screws.A ria was used in the tibia to retrieve bone graft.Surgeon corrected the deformity, and patient¿s tibia was revised with a tibial nail advanced, and placement of new screws in remaining two plates.Fragment were generated and was removed easily.The procedure was successfully completed with five minutes surgical delay.This complaint involves (42) devices.This report is for one 3.5mm lcp plate 12 holes 163mm this report is for 12 of the 42 devices.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: d4: lot part number : 223.621 lot number: 5752383 part manufacture date: april 11, 2008 dhr record review: a review of the device history record revealed no complaint related anomalies.The devices history record show this lot of 3.5mm lcp plate 12 holes product was processed through the normal manufacturing and inspection operations with no rework nor nonconformities noted.This lot met all dimensional, visual, and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.A review of the raw material device history record revealed this raw material lot had one material review record, issued against it.Raw material lot 4969084 had 65 pieces rejected and scrapped for cosmetic - visual defects for scratches and scuff marks.This nonconformance does not contribute to the complaint condition.There were no other nonconformances documented against this raw material lot at the time of manufacture.Photo investigation: the part was not returned for investigation.A photo investigation was carried out using the x-rays.The part was removed from the patient body due to a non-union of the left humerus and right tibia.There appears to be no issues with the part from the x-rays provided.A root cause of the issue cannot be determined from the provided information.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.Conclusion: the complaint condition cannot be confirmed based on the image provided.During the investigation, no product design issues, or discrepancies were observed.No manufacturing issues were noted during investigation.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Therefore, it has been determined that no corrective and/or preventive action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|