Model Number URF-P6
Device Problems Display or Visual Feedback Problem (1184); Poor Quality Image (1408)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The returned device was evaluated.Device inspection found the reported issue was able to be duplicated.Inspection found the device scope image was observed with 40 ig (image guide) breakage.The bending section was found to be broken, leaking at the ¿a-rubber¿ was determined and cracked glue was observed.Based on evaluation findings the failures found could be attributed to use handling and or maintenance issue.This report will be supplemented accordingly following investigations.
 
Event Description
It was reported that the scope image has broken fibers.The issue occurred during preparation for use.There was no patient involvement, no user injury reported due to the event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found the device is a refurbished.The subject device has not been repaired within the past year.Based on the results of the investigation, the cause of the suggested event cannot be specified.From the evaluation findings, it was presumed that the user manipulated the device and excessive stress applied to the bending section, which damaged the bending tube: user can properly detect the suggested event by handling device in accordance with the following ifu (instruction for use) description: precautions : perform a leakage test on the endoscope after each precleaning procedure.Do not use the endoscope if a leak is detected.Use of an endoscope with a leak may cause a sudden loss of the endoscopic image, damage to the bending mechanism, or other malfunctions.Use of a leaking endoscope may also pose an infection control risk.·inspection of the endoscope : inspect the external surface of the entire insertion section, including the bending section and the distal end for dents, bulges, swelling, scratches, peeling of coating, holes, sagging, transformation, bends, adhesion of foreign bodies, missing parts, protruding objects, or other irregularities.¿ ·do not operate the angulation control lever with excessive force in a narrow space to the opposite direction from the bending direction while the distal end of the endoscope is not moved.The bending section may be damaged.Check the tip position of the endoscope and the shape of the bending section using fluoroscopy, etc.Do not insert the insertion tube with excessive force and twist.·do not insert the insertion tube with excessive force into the ureter or calix.The bending section may be damaged.Olympus will continue to monitor complaints for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO FIBERSCOPE
Type of Device
URETERO-RENO FIBERSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key11992590
MDR Text Key256078138
Report Number8010047-2021-07493
Device Sequence Number1
Product Code FGB
UDI-Device Identifier04953170340802
UDI-Public04953170340802
Combination Product (y/n)N
PMA/PMN Number
K172298
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 10/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/14/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-P6
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/11/2021
Was the Report Sent to FDA? No
Date Manufacturer Received10/20/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-